Calibrator Setup
Unpack and Inspect
Unpack the instrument carefully and examine it for any damage that could have occurred
during shipment. If there is shipping damage, notify the Fluke Calibration and the carrier
immediately. Table 2-6 lists the equipment and the accessories that comes with the
Calibrator. Verify that all the equipment and accessories in Table 2-6 are in the box.
Table 2-6. Parts and Accessories
Name Quantity
9190-INSX Insert (X=A, B, C, D, E, or F) 1
2-meter (6-foot) Mains Power Cord 1
USB Cable 1
Getting Started Manual 1
Product CD that contains manuals and remote interface driver files 1
9930 Interface-it Calibration Software and Users Guide 1
Report of Calibration and Calibration Label 1
Well Insulator Cap 1
Insert Removal Tool 1
Clamp-on Ferrites (-P model only) 4
6-pin DIN Connector (-P model only) 1
Test Lead Kit (-P model only) 1
Placement
Put the Calibrator on a clean, flat surface. Make sure the Calibrator is 150 mm (6 inches)
away from all objects. For best results, choose a location to set up the Product where
room temperature changes are minimum.
Warning
To prevent possible fire or personal injury:
Do not operate Product in orientations other than upright. A
fire hazard can be made if the Product is put on its side.
Do not remove Inserts when the Product shows
temperatures more than 50 °C.
Do not operate near flammable materials.
Do not touch the well access surface of the Product.
Do not turn off the Product when the temperature is above
100 °C. Set a SETPOINT temperature below 100 °C and let
the Product cool.