9190A
Operators Manual
3-6
USB Remote Interface
Use the instructions in the subsequent sections to wire and configure a USB Remote
Interface.
Wiring
The Calibrator comes with a standard USB cable. The USB cable has a 1394 connector
(Calibrator connection) on one end and a standard USB connector on the other (computer
connection).
To use the USB interface for remote operation, a driver needs to be installed on the
computer. After the driver is installed and the Calibrator is connected to the computer, the
USB connection appears as a virtual COM port in the PC communication software (for
example, MS Hyper-terminal). The driver is included on the CD that shipped with the
Calibrator. If the CD is lost, or a updated driver is necessary, go to
http://www.ftdichip.com/Drivers/VCP.thm.
Communication Setup
To use the USB remote interface, the Calibrator must be configured to communicate with
the computer. Configuration is done in the Communication Setup menu on the Calibrator.
Table 3-3. Communication Setup Menu
66.03°C
COMMUNICATION SETUP
BAUD RATE: 9600
LINEFEED: ON
COMM
SETUP
SYSTEM
MENU
SYSTEM
SETUP
MENU
1
2
gzs016.eps
Item Function
BAUD RATE is the serial communication transmission rate or baud rate. BAUD may be set to
1200, 2400, 4800, 9600, 19200, or 38400 baud.
LINEFEED turns on or turns off transmission of a line feed character (LF, ASCII 10) after
transmission of any carriage-return.
Setup communications as follows:
1. Push .
2. Push .
3. Push .
4. Push .
5. Highlight
BAUD RATE and push .
6. Push or to change the
BAUD RATE (see item in Table 3-2).
Note
Set the BAUD RATE to the baud rate that the computer serial port is configured for.
7. Push to save.