SOFT CUTOUT:
+090.00 °C
(HARD CUTOUT: +160.00 °C)
gzs037.eps
Figure 2-13. Soft Cutout
Reset an Over-Temperature Cutout
If the Soft or Hard Cutout is exceeded, the Calibrator enters Cutout Mode and must be
reset.
To exit Cutout mode and reset the Calibrator:
1. Let the Calibrator cool to below the Soft Cutout limit.
2. After cooled, push followed by to reset the Calibrator.
Programs (Automated Tests)
The Program menu (PROG MENU) supplies access to the automated and manual
Program selections (see “Program Setup” on page 2-15). A Program is an automated test
that can be configured and run on an instrument.
Run a Program
To run a Program:
1. Push .
2. Push .
3. Push .
4. Push to change the
TEST STATUS to RUN.
5. Push to save.
Stop a Program
To stop a Program:
1. Push .
2. Push .
3. Push .
4. Push to change the
TEST STATUS to OFF.
5. Push to save.