EasyManua.ls Logo

Fluke 9190A - RS-232 Remote Serial Interface; Wiring

Fluke 9190A
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9190A
Operators Manual
3-4
RS-232 Remote Serial Interface
Use the instructions in the subsequent sections to wire and configure a RS-232 Remote
Serial Interface.
Wiring
Wire the RS-232 Remote Serial Interface connection as follows:
1. Wire the remote serial interface cable as shown in Figure 3-1.
Note
To prevent noise on the signal wires, use a shielded, low resistance cable
between the Product and PC.
Instrument
Connector
(DB 9-Pin)
1 NC
2 RxD
3 TxD
4 ND
5 GND
6 NC
7 RTS
8 CTS
9 NC
Computer (DTE)
Connector
(DB 9-Pin)
1 DCD
2 RxD
3 TxD
4 DTR
5 GND
6 DSR
7 RTS
8 CTS
9 NC
Instrument
Connector
(DB 9-Pin)
1 NC
2 RxD
3 TxD
4 ND
5 GND
6 NC
7 RTS
8 CTS
9 NC
Computer (DTE)
Connector
(DB 25-Pin)
1 DCD
2 RxD
3 TxD
4 DTR
5 GND
6 DSR
7 RTS
8 CTS
9 NC
gzs048.eps
Figure 3-1. RS-232
2. Connect the DB-9 connector wired for the instrument to the 9-pin Subminiature
Serial Connector (RS-232) on the Power and Remote Interface Panel (see Table 3-1).
3. Connect the DB-9 or DB-25 connector wired for the computer to the 9-pin or 25-pin
serial port on the computer.