Remote Operation
Remote Operation Commands 3
3-25
SOUR:STAB:LIM[?] [n]
Read or set the control temperature stability limit, °C or °F. “n” is a positive real value.
Range = {0.01 to 9.99 (°C)}; Default: 0.05 (°C).
Read Example: SOUR:STAB:LIM?
Response: 0.05
Set Example: SOUR:STAB:LIM 0.03
SOUR:STAB:TEST?
Read the temperature stability test results. A value of 0 is returned if the controller is not stable at
the current SETPOINT. Otherwise a value of 1 is returned if the controller is stable at the current
SETPOINT.
Stable = 1; Unstable = 0.
Example: SOUR:STAB:TEST?
Response: 0
SYST:BEEP:IMM
Beep the system beeper. The system beeper should make an audible sound in response to this
command.
Example: SYST:BEEP:IMM
SYST:BEEP:KEYB[?] [n]
Read or set the keyboard beep function.
0=Off, 1=On. Default: 1.
Read Example: SYST:BEEP:KEYB?
Response: 1
Set Example: SYST:BEEP:KEYB 1
SYST:CODE:LANG?
Read the language set option: 1: European; 2: Asian.
European: ENGLish (default), FRENch, GERMan, PORTuguese, SPANish, RUSSian
Asian: ENGLish (default), CHINese, JAPanese, KORean
Example: SYST:CODE:LANG?
Response: 2
SYST:CODE:VERS?
Read the main code version. Supplies the user with the version of the main processor code.
Example: SYST:CODE:VERS?
Response: 1.10
SYST:COMM:SER:BAUD[?] [<baud>]
Read or set serial interface baud rate where “baud” is a standard baud rate value.
Range baud = {1200, 2400, 4800, 9600, 19200, and 38400}; Default: 9600.
Read Example: SYST:COMM:SER:BAUD? Response: 2400
Set Example: SYST:COMM:SER:BAUD 9600