Calibration
Temperature Source Calibration 4
4-7
9190A
Secondary PRT (30 mm)
Test Sleeve Insert
Primary PRT (0 mm)
To Readout (see “Suggested Equipment” in Table 4-2)
gzs052.eps
Figure 4-2. Test Sleeve Orientation
Connect both PRTs to the readout. It is best to make simultaneous measurements of both
the primary and secondary PRT when PRT data is measured from the readout. If the
readout only has one input channel, take the primary PRT readings back to back with the
secondary PRT without delay.
A remote connection to the product is required to complete this procedure. Use the
instructions in Chapter 3 “Remote Operation” to set up the remote connection to the
UUT.