Technical Data
General Specifications
7
3. Technical Data
3.1. General Specifications
General Specifications
Environmental Rating for housing
IP65 (IEC529) / NEMA 4
Ambient Temp. without cooling
All models, except E2RL
E2RL
0 - 65°C (32 - 149°F)
0 - 60°C (32 - 140°F)
Ambient Temp. with air cooling
0 - 120°C (32 - 248°F)
Ambient Temp. with water cooling
0 - 175°C (32 - 347°F)
Ambient Temp. with ThermoJacket
0 - 315°C (32 - 600°F)
Storage Temperature
-20 to 70°C (-4 to 158°F)
Relative Humidity
10 to 95%, non-condensing
at 22°C to 43°C (72°F - 110°F)
EMC
EN 61326-1:2013
Safety
EN 60825-1:2008-05
FDA laser safety compliant
Mechanical Shock
IEC 68-2-27 (5 G, 11 msec duration, 3 axes)
Vibration
IEC 68-2-6 (2 G, 10 to 150 Hz, 3 axes)
Warm up Period
15 minutes
Weight
Endurance sensor
Air / Water cooled housing
Mounting nut
Fixed mounting bracket
1220g (2.69 lbs)
1760g (3.88 lbs)
62g (0.14 lbs)
264g (0.58 lbs)
Sensor Head Housing Material
Stainless Steel
Mat.-No.: 1.4305, Mat.-Name.: X8CrNiS18-9
Control Panel (User Interface)
Upper Display (green): 7-segment, 4 digits LED type to display the measured object
temperature or error codes.
Lower Display (green/red): Background illuminated graphics display type. Resolution
is 32 * 136 pixels to display 2 text lines of about 16
characters per line. It is the main screen/menu display,
which shows all information and configuration topics.
LED1 (red): Indicates the Laser, LED or Video on/off status
LED2 (green): Sensor alarm status, steady green after warm-up phase
Control buttons: 4 pushbuttons, to step through the menu and to enter setup values