Endurance® Series
Users Manual
64
9.1.1. High Temp. Multi-conductor Cable (E-2CCBxx)
Use the High Temp. 12-wire multi conductor cable (E-2CCBxx) for wiring the Endurance sensor
with the 24 VDC power supply, all inputs, outputs, and the RS485 interface. It is a shielded 12-
conductor cable, made of 2 twisted pairs plus 8 separate wires, equipped with a M16 DIN
connector on one side and wire sleeves at the counter side. The cable is Teflon coated and
withstands ambient temperatures from -80 to 200°C (-112°F to 392°F). Teflon coated
temperature cables have well to excellent resistance to oxidation, heat, weather, sun, ozone,
flame, water, acid, alkalis, and alcohol, but poor resistance to gasoline, kerosene, and
degreaser solvents.
Purchasable High Temp. 12-wire multi-conductor cables lengths are 4 m (13 ft.), 8 m (26 ft.),
15 m (50 ft.), 30 m (100ft.), 60 m (200 ft.), see Table 5.
Temperature: UL-rated at -80 to 200°C (-112°F to 392°F)
Cable material Teflon
Cable diameter: 7 mm (0.275 in) nominal
Conductors:
Power supply 2 wires (black/red)
Conductor: 0.3 mm² (AWG 22), stranded tinned copper
Isolation: FEP 0.15 mm wall (0.006 in)
Shield: none
RS485 interface 2 twisted pairs (black/white and purple/gray)
Conductor: 0,22 mm² (AWG 24), stranded tinned copper
Isolation: FEP 0.15 mm wall (0.006 in)
Shield: Aluminized Mylar with drain wire
Outputs and Ground 6 wires (green/brown/blue/orange/yellow/clear)
Conductor: 0,22 mm² (AWG 24), stranded tinned copper
Isolation: FEP 0.15 mm wall (0.006 in)
Shield: none
Teflon develops poisonous gasses, when it is exposed to flames!
If you cut the cable to shorten it, notice that both sets of twisted-pair wires
have drain wires inside their insulation. These drain wires (and the white wire
that is not part of the twisted pair) must be connected to the terminal labeled
CLEAR.
If you purchase your own High Temp. Multi-conductor cable, use wire with the
same specifications as herein mentioned. Maximum RS485 cable length is
1.200 m (4000 ft). Power supply (24 VDC) feed in distance to the Endurance
sensor should not extend the 60 m (200 ft.) limit.
An ordered Multi-Conductor Cable does not include a terminal block!
Figure 47: High Temp. Multi-Conductor Cable with M16 Connector (E-2CCBxx)