There is no warranty repair possible in case of condensation within the
housing!
To avoid condensation, the temperature of the cooling media and the flow rate must be selected
to ensure a minimum device temperature. The minimum sensor temperature depends on the
ambient temperature and the relative humidity. Please consider the following table.
Table 4: Minimum device temperatures [°C/°F]
Ambient Temperature [°C/°F]
Temperatures higher than 60°C (140°F) for the E2RL sensor
model or 65°C (149°F) for the other model variants are not
recommended due to the temperature limitation of the sensor.
Example:
Ambient temperature = 50 °C
Relative humidity = 40 %
Minimum device temperature = 30 °C
The use of lower temperatures is at your own risk!