Programming Guide
Command List
87
Table 8: Assignment of Error-Codes
2
15
2
14
2
13
2
12
2
11
2
10
2
9
2
8
2
7
2
6
2
5
2
4
2
3
2
2
2
1
2
0
Alarm detection
Narrow band temperature over range
Narrow band temperature under range
Wide band temperature over range
Wide band temperature under range
Two-color temperature over range
Two-color temperature under range
Attenuation > 95% ("dirty window")** (1)
Attenuation too high (> 95%) (1)
Energy too low
Narrow band detector failure
Wide band detector failure
Internal temperature under range
Internal temperature over range
Heater control temperature under range
Heater control temperature over range
High temperature limit
XH
n.n nnnn.n
0.09999.0 (5)
Set at factory
calibration
Sensor initialization
XI
n
0 = flag reset, 1 = flag set
1
Laser / LED / Video
switching
XL
n
0 = off
1 = on
2 = flashing
3 = switching via external
trigger edge
4 = switching via external
trigger level
0 = off
Sensor model type
XM
X
L = Low Temp., H = Hi
Temp
Set at factory
calibration
0 - 20 mA / 4-20 mA
analog output
XO
n
0 = 0 - 20 mA, 4 = 4 - 20
mA
4
Sensor firmware revision
no.
XR
Xn
e.g. 1.02.11
Set at factory
calibration
Sensor analog part
revision no.
XRA
Xn
e.g. 1.02.01
Set at factory
calibration
Setpoint / Relay function
XS
n.n nnnn.n
0.0 to 3200.0°C /
5792.0°F (8)
0.0
Trigger
XT
N
0 = inactive, 1 = active
0
Identify unit
XU
Varies
e.g. E1RL-F2-V-0-0
Set at factory
calibration
Sensor serial number
XV
nnnnnnnn
V
e.g. 31712345 (8 digits)
Set at factory
calibration
Attenuation to activate
relay
Y
nn
0 to 95% energy
95%
Attenuation for failsafe
Z
nn
0 to 99% energy
reduction
95%
Notes:
(1) Commands may appear as Polled for (queried), Burst string item or Set command
(2) n = number, X = uppercase letter.
(3) see section 10.3.2 Burst Mode, page 84
(4) Setting either Average, Peak Hold or Valley Hold, sets non concerned signal post processing settings to
factory default value
(5) In current scale °C or °F
(6) The sensor restarts after a baud rate change. (Command is not allowed in multidrop mode.)
(7) No effect if relay in alarm mode.
(8) Non-zero setpoint value puts unit in setpoint mode. Setpoint is in current scale °C or °F and must be within
unit’s temperature range.
(9) Error Codes returned out of ?EC-Command (16 Bit-Word, 0000000000000000 1111111111111111)