Endurance® Series
Users Manual
86
Analog input mode
INM
n
0 = 0-20mA, 4 = 4-20mA
4
IP Address
IP
nnn.nnn.nnn.nnn
0.0.0.1 - 255.255.255.255
192.168.42.132
Switch panel lock
J
X
L = Locked
U = Unlocked
Unlocked
Relay alarm output
control
K
n
0 = Permanently Open
2
1 = Permanently Closed
2 = Normally Open
3 = Normally Closed
Bottom of mA
temperature range
L
n.n nnnn.n
0.0 9999.0 (°C or °F)
Lower end of
sensor range
ModeER series
M
n
1 = 1 - color
2
2 = 2 - color
MAC Hardware Address
MAC
nnnnnnnnnnnn
e.g. 001d8d2aaa01
Set at factory
calibration
Target temp:1C narrow
N
n.n - nnnn.n
Net Mask
NM
nnn.nnn.nnn.nnn
0.0.0.1 - 255.255.255.255
255.255.255.0
Output current
O
nn
00 = controlled by unit
00
02 = under range
21 = over range
00 20 = current in mA
Output format
OIF
0 = temperature value
format is float with one
decimal place
1 = temperature value
format is integer with four
digits
0
Peak hold time (4)
P
n.n - nnn.n
0.0 300.0 sec (300 s =
)
0.0
IP Portaddress
PORT
n - nnnnn
1 - 65535
6363
Wide Power
Q
n.nnnnnnn
Narrow power
R
n.nnnnnnn
Video relative reticle
diameter
RC
n.n nn.nn
Video relative reticle X-
position
RX
n.n nn.nn
Video relative reticle Y-
position
RY
n.n nn.nn
Slope
S
n.nnn
0.850 1.150
1.000
Alarm source
SAS
0 = object temperature
1 = internal temperature
0
Slope source
SS
X
I or E
I
Set target temperature
STT
n.n nnnn.n
0.0 9999.0 (5)
Set at factory
calibration
Target temp: 2C
T
n.n - nnnn.n
Terminator resistor
TR
n
0 = OFF, 1 = ON
0 = OFF
TCP/IP time out interval
TTI
n - nnn
0 = , 1 240 sec
0
Temperature units
(scale)
U
X
C or F
non-US: C
Poll/burst mode
V
X
B = Burst , P = Polled
P = Polled
Target temp:1C wide
W
n.n - nnnn.n
(5)
Web server ON/OFF
WS
n
0 = OFF, 1 = ON
0 = OFF
Burst string contents (3)
X$
Multidrop address
XA
nnn
000 to 032
000
Low temperature limit
XB
n.n - nnnn.n
0.09999.0 (5)
Set at factory
calibration
Deadband (7)
XD
nn
01 55 in °C / 01 99 in
°F
02
Restore factory defaults
XF
Transmissivity
XG
n.nn
0.10 1.10
1.00