EasyManua.ls Logo

Fluke Endurance E3ML - Command List

Fluke Endurance E3ML
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Guide
Command List
85
10.4. Command List
The table below describes the available commands via RS485 or LAN/Ethernet interface.
Table 7: Command List
Description
Char
Format
(2)
P
(1)
B
(1)
S
(1)
Legal Values
Factory
Default
Burst string format
$
(3)
(3)
UTSI
Show list of commands
?
Ambient correction
A
min/max range
Low end of
sensor range
Advanced hold w.
average
AA
0.0-300.0s
000.0
Ambient compens.
control
AC
0, 1 or 2
0
Top of mA range
AH
min/max range
High end of
sensor range
Bottom of mA range
AL
min/max range
Low end of
sensor range
Measured attenuation
B
nn - nn
00 to 99%
n/a
Burst speed
BS
n - nnnnn
5 - 10000msec
32msec
Advanced hold threshold
C
min/max range
Low end of
sensor range
Current emissivity
CE
n.nnn n.nnn
0.100 1.100
1.000
Camera color mode
CCM
C = displays all colors
M = converts all colors
into grayscale
R = converts red colors
into grayscale
G = converts green colors
into grayscale
B = converts blue colors
into grayscale
C
Camera auto gain mode
CGM
0 = off
1 = on
1
Baud rate (6)
D
nnn - nnnn
12 = 1200 baud
38400 baud
24 = 2400 baud
96 = 9600 baud
192 = 19200 baud
384 = 38400 baud
576 = 57600 baud
1152 = 115200 baud
Digital filter
DF
n
0 = OFF, 1 = ON
1
DHCP / BOOTP
DHCP
n
0 = OFF, 1 = DHCP ON
2 = BOOTP ON
0
Sensor gain
DG
n.nnnnnn -
n.nnnnnn
0.800000 up to 1.200000
1.000000
Sensor offset
DO
-nnn - +nnn
-200 up to +200
0
Emissivity
E
n.nnn
0.100 1.100
1.000
Extension board
temperature
EBT
n.n - nnn.n
0.0 999.0 (°C or °F)
Error Codes (9)
EC
nnnnnnnn
0000 FFFF (Hex)
Emissivity source
ES
X
I or E
I
Valley hold time (4)
F
n.n - nnn.n
0.0 300.0 sec (300 s =
)
000.0
Average time (4)
G
n.n - nnn.n
0.0 300.0 sec (300 s =
)
000.0
Gateway Address
GW
nnn.nnn.nnn.nnn
0.0.0.0 - 255.255.255.255
192.168.42.1
Top of mA temperature
range
H
nnnn.n nnnn.n
min/max range (°C or °F)
Upper end of
sensor range
Sensor internal ambient
I
n.n - nnn.n
0°C/32°F 65°C/149°F
Analog input mA
IN
nn.nn nn.nn
0-20 or 4-20
w ww. . com
information@itm.com1.800.561.8187

Table of Contents

Related product manuals