Signal Processing
Factory Defaults
45
7.9. Factory Defaults
To globally reset the unit to its factory default settings, go to the “factory default” menu item
under the configuration screen menu display. The baud rate and communications mode (single
device or multiple devices / multidrop) will not be affected.
Table 3: Factory Defaults
Parameter
E1M, E2M, E3M
E1R, E2R
Mode (1C / 2C)
1C
2C
Temperature Unit (°C / °F)
°C
°C
Slope
n/a
1.000
Emissivity
1.000
1.000
Transmissivity
1.00
1.00
Average
0.0
0.0
Peak Hold
0.0
0.0
Valley Hold
0.0
0.0
SETPOINT in (°C / °F)
0.0
0.0
DEADBAND in (°C / °F)
2
2
RS485 Communication Mode
2-wire , 38.400 Baud *
2-wire, 38.400 Baud *
MULTIDROP ADDRESS
000 (single sensor)
000 (single sensor)
TERMINAL RESISTOR
OFF
OFF
ETHERNET DHCP
OFF
OFF
ETHERNET IP-ADDRESS
192.168.42.132
192.168.42.132
ETHERNET NETMASK
255.255.255.0
255.255.255.0
ETHERNET GATEWAY ADDR.
192.168.42.1
192.168.42.1
ETHERNET PORTNUMBER
6363
6363
WEB SERVER
OFF
OFF
ANALOG OUTPUT MODE
4 20mA
4 20mA
OUT Lo LIMIT for 4 mA
Low limit sensor temp.
(e.g. 400.0°C)
Low limit sensor temp.
(e.g. 400.0°C)
OUT Hi LIMIT for 20 mA
High limit sensor temp.
(e.g.1800.0°C)
High limit sensor temp.
(e.g.1800.0°C)
ANALOG INPUT MODE
4 20mA
4 20mA
IN Lo LIMIT for 4 mA
Low limit sensor temp.
(e.g. 400.0°C)
Low limit sensor temp.
(e.g. 400.0°C)
IN Hi LIMIT for 20 mA
High limit sensor temp.
(e.g.1800.0°C)
High limit sensor temp.
(e.g.1800.0°C)
Serial Output Transmission Mode
Burst mode,
Default string = UTSI
Burst mode,
Default string = UTSI
Relay Output Control
Controlled by unit,
NO function,
indicates failsafe alarms
Controlled by unit,
NO function,
indicates failsafe alarms
Set Output Current
Controlled by unit,
4-20 mA
Controlled by unit,
4-20 mA
Lockout Control Panel Access
Unlocked
Unlocked
* RS485 Modes, like Baud Rate or 2-wire half duplex are unchanged, when the factory
defaults are restored
w ww. . com
information@itm.com1.800.561.8187