Sensor Location
Sensor Placement (2-Color Mode)
17
4.6. Sensor Placement (2-Color Mode)
The following figure demonstrates the sensor placement under various conditions, where valid
2-Color temperature measurements are possible. Note, however, that if the sensor signal is
reduced more than 95% (including emissivity and obscuration of the target), the sensor
accuracy also degrades.
Figure 5: Sensor Placement in 2-Color Mode
Sighting hole smaller
than the sensor’s field of
view
Emitted
energy
Dirty lens or
dirty sighting window
Emitted
energy
Smoke, steam, dust,
gas in atmosphere
Emitted
energy
Target smaller than field of view
and / or moves or vibrates in and
out of field of view (e.g. wire)
Emitted
energy