Maintenance
Troubleshooting Minor Problems
89
11. Maintenance
Our sales representatives and customer service are always at your disposal for questions
regarding application assistance, calibration, repair, and solutions to specific problems. Please
contact your local sales representative if you need assistance. In many cases, problems can
be solved over the telephone. If you need to return equipment for servicing, calibration, or
repair, please contact our Service Department before shipping. Phone numbers are listed at
the beginning of this document.
11.1. Troubleshooting Minor Problems
Table 10: Troubleshooting
Symptom
Probable Cause
Solution
No output
No power to instrument
Check the power supply
Erroneous temperature
Faulty sensor cable
Verify cable continuity
Erroneous temperature
Field of view obstruction
Remove the obstruction
Erroneous temperature
Window lens
Clean the lens
Erroneous temperature
Wrong slope or emissivity
Correct the setting
Temperature fluctuates
Wrong signal processing
Correct Peak Hold or Average settings
11.2. Fail-Safe Operation
The Fail-Safe system is designed to alert the operator and provide a safe output in case of any
system failure. Basically, it is designed to shut down the process in the event of a set-up error,
system error, or a failure in the sensor electronics.
Warning
The Fail-Safe circuit should never be relied on exclusively to protect critical
heating processes. Other safety devices should also be used to supplement
this function!