Technical Data
Optical Specifications
13
3.4.1. Measurement spot size regarding the selected focus and model options
The properties of Endurance devices with an integrated sensor head are:
Variable focus
Through-the-lens sighting/pointing option (LED, LASER, Video-Camera)
Parallax-free optics
May be mounted almost everywhere
Three variable focus distances are optional orderable:
F0 (Narrow Focus) models can be focused from 190 to 300mm (7.5 - 12'')
F1 (Close Focus) models can be focused from 300 to 600 mm (12 - 24'')
F2 (Standard Focus) models can be focused from 600 to infinity (24'' - infinity)
For measurements performed in the 1-Color mode, make sure that the target completely fills
the measurement spot. The target spot size for a properly focused target with the given distance
to the target can be determined by using the following formula.
Divide the distance (D) in the picture below by the (D:S) specification to get the target
spot size.
Example for an E1ML device (D:S = 160:1) under different focus options:
F0: 190 300mm The measurement spot diameter varies from 1.1875 1.875mm
F1: 300 600mm The measurement spot diameter varies from 1.875 3.75mm
F2: 600 3000mm The measurement spot diameter varies from 3.75 18.75mm
Figure 2: Spot size calculation regarding the distance to the integrated sensor head
The spot size calculated using this method is valid only at the focus distance.
Spot sizes out of focus distances will vary from the rule.
D = Distance
S = Spot size
w ww. . com