8.8.3.4. Pyrometer Diagnostics
The Ethernet/IP device has a designated status register. The bits of this register make up for
an error code, which is sent as a part of input data.
0x05
Status
DWORD
4 Byte
and which can be translated using the table below.
Bit
Description
0
Heater temperature over range
1
Heater temperature under range
2
Internal temperature over range
3
Internal temperature under range
4
Wide band detector failure
5
Narrow band detector failure
6
Energy too low
7
Attenuation for failsafe too high
8
Attenuation to activate relay too high
9
Two color temperature under range
10
Two color temperature over range
11
Wide band temperature under range
12
Wide band temperature over range
13
Narrow band temperature under range
14
Narrow band temperature over range
15
Alarm
16
Video overflow
17
EthernetIP not ready
18
Heater not ready