EasyManua.ls Logo

Fronius Calibration system 2.0 - Page 79

Fronius Calibration system 2.0
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Notes on voltage calibration:
-
The maximum voltage during calibration is 30 V.
-
The same resistance of 480 mOhm is used for all measuring points.
-
The calibration software automatically specifies the measuring points to be
measured.
-
The voltage is regulated by the current;
the predefined current values are standard values and may have to be adjus-
ted to achieve the specified voltages.
Recommended points for calibration:
Voltage [V] Resistance [mOhm] Current on the device
[A]
10 480 20
15 480 29
20 480 39
25 480 49
30 480 59
Calibration procedure
1
Start the calibration program
2
Click [New calibration]
The preliminary settings / contractor data are displayed.
3
Enter the details for a new contractor (= Person performing calibration):
-
Name
-
Address
-
Postcode
-
Calibration technician
4
Click on [Add]
or
select current contractor from the list shown
5
Enter the serial numbers of the measuring equipment:
-
Calibration system
-
Gas measurement system
6
Click [Next >]
The preliminary settings / customer data are displayed.
7
Enter the details for a new client:
-
Name
-
Address
-
Postcode
-
Customer number
79
EN

Table of Contents