EasyManuals Logo

GE Corometrics 250 Series User Manual

GE Corometrics 250 Series
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
4-54 250 Series Maternal/Fetal Monitor Revision D
2020551-001
Maintenance: Power Supply Voltages—Verification
Isolated Power Supply Board Voltages
Verify the following voltages, being sure to use the correct ground reference points.
Isolated FECG/UA Board Voltages
This procedure performs an adjustment on the FECG/UA Board.
Connect the positive lead of a digital voltmeter to TP1 on the FECG/UA Board.
Connect the negative lead to TP2 or TP3 (isolated ground).
Adjust R28 for a reading of +4.00 Vdc ± 0.01 Vdc.
Table 20. Main Board Power Supply Voltages
2023111-001 Signal Name Voltage Level
Pin 1 +12EL +12 Vdc ± 0.5 Vdc
Pin 2 +20I +20 Vdc ± 0.5 Vdc
Pin 3 +15BP +15 Vdc ± 0.5 Vdc
Pin 4 15V –15 Vdc ± 0.5 Vdc
Pin 5 +15V +15 Vdc ± 0.5 Vdc
Pin 6 +12A +12 Vdc ± 0.5 Vdc
Pin 7 +5V +5 Vdc ± 0.5 Vdc
Pin 8 GND
Pin 9 No Connection
Pin 10 Keying
Ground Reference Test Point Measurement Limit
TP8 TP7 16.5 ±.15 Vdc
TP8 TP17 15 ±.75 Vdc
TP8 TP16 -15 ±.75 Vdc
TP8 TP18 5 ±.25 Vdc
TP5 TP10 15 ±.75 Vdc
TP5 TP11 -15 ±.75 Vdc

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals