Revision D 250 Series Maternal/Fetal Monitor 3-21
2020551-001
Installation: Communications Setup Screen
Communications Setup Screen
The Communications Setup screen, allows you to configure the unit’s
J109, J110, and
J111 RS-232C
Serial Interface ports for connecting to optional peripheral equipment.
Each port is configured for baudrate and mode.
To use the Communications Setup screen:
Select the Service softkey from the General Setup screen.
The Service Lock screen displays.
Enter the correct access code; Install Options Screen 1 displays.
Select the COMM softkey on Install Options Screen 1.
The Communications Setup screen appears.
Baudrate
This field selects the baudrate for communication with an external device. Use the
Trim Knob control to cycle through the available settings: 600, 1200, 2400, 4800,
9600, and 19,200 bps. (Each port is factory set to 2400 bps.) Refer to “Peripheral
Components” on page 2-14 for more information about configuring this field.
Mode
This field selects the mode for communication with an external device. Use the Trim
Knob control to cycle through the available settings: Nellcor, Factory, Critikon, 1371,
1371/NOTES, LOOPBACK, 115, and 115 X/R. (Each port is factory set to the Nellcor
mode.) Refer to “Peripheral Components” on page 2-14, for more information about
configuring this field.
NOTE: At the factory default settings of 2400 bps and the Nellcor mode, connectors
J109 and J111 are ready for connection to a Nellcor Puritan Bennett (NPB)
Model N-200 or Model N-400. Be advised that connector J110 does not
support a connection to a NPB monitor; therefore, this connector will
have to be configured for use with another device, if appropriate.