Revision D 250 Series Maternal/Fetal Monitor 3-23
2020551-001
Installation: Factory Defaults
Factory Defaults
Table 1. Summary of Factory Defaults
Setup
Screen
Field Description Factory Default Default Options Hospital Preference
FECG or
US/US2
FHR Volume 5 0-9
FHR Alarm Limits High Low
160 120 bpm
200-140, Off
60-140, Off
Audio Alarms On On, Off
Volume 5 1-9
NIBP Initial Target Pressure 135 mmHg
(18.0 kPa)
100-250 mmHg in increments of
5;
(13.3-33.3 kPa in 0.7 steps
Mode Manual Manual, 1, 2, 3, 4, 5, 10, 15, 20,
30, 40, 45, 60, 90, 120 min
NIBP Done Vol 5 0-9
Alarm (mmHg mode)
Systolic
Diastolic
MAP
MHR/P
High Low
160 90 mmHg
90 50 mmHg
140 50 mmHg
120 50 bpm
High Low
70-240 50-150
70-130 30-120
70-150 30-120
100-250 35-120
Alarm (kPa mode)
Systolic
Diastolic
MAP
MHR/P
High Low
21.3 12.0 kPa
12.0 6.7 kPa
18.7 6.7 kPa
120 50 bpm
High Low
9.3-32.0 6.7-20.0
9.3-17.3 4.0-16.0
9.3-20.0 4.0-16.
100-250 35-120
Volume 5 1-9