3-24 250 Series Maternal/Fetal Monitor Revision D
2020551-001
Installation: Factory Defaults
MSpO
2
Response Time
(Nellcor 506)
Fast Normal, Fast
Response Time
(Nellcor NELL-3)
Fast Fast
Averaging Time (Masimo) 8 seconds 2, 4, 8, 10, 12, 14, 16 seconds
Sensitivity (Masimo) Normal Normal, Maximum
Print Interval 5 minutes Off, 2, 5, 10, 15, 30, 60 minutes
%O
2
Trace Off On, Off
Alarms
MSpO
2
MHR/P
High Low
100 95%
120 50 bpm
High Low
85-100 80-99
100-250 35-120
Volume 5 1-9
MHR/P Source Auto Auto, MECG, MSp0
2
, NIBP
MHR/P Trace Off On, Off
Volume 0 0-9
Alarms High Low
120 50 bpm
High Low
100-250 35-120
Alarm Volume 5 1-9
MECG Lead II I, II, III
Pacer Off On, Off
Normal
Operation
(Waveform Display) MECG FECG, MECG, MSp0
2
, Off
(MECG Waveform Size) 1X (1 mV/cm) 0.25X, 0.5X, 1X, 2X, 4X, Auto
Master
Alarm
Setup
Alarm Limits (mmHg)
Systolic
Diastolic
MAP
MHR/P
MSpO
2
High Low
160 90mmHg
90 50 mmHg
140 50 mmHg
120 50 bpm
100 95%
High Low
70-240 50-150
70-130 30-120
70-150 30-120
100-25 35-120
85-100 80-99
Alarm (kPa mode)
Systolic
Diastolic
MAP
MHR/P
High Low
21.3 12.0 kPa
12.0 6.7 kPa
18.7 6.7 kPa
120 50 bpm
High Low
9.3-32.0 6.7-20.0
9.3-17.3 4.0-16.0
9.3-20.0 4.0-16.
100-250 35-120
Volume 5 1-9
Table 1. Summary of Factory Defaults
Setup
Screen
Field Description Factory Default Default Options Hospital Preference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals