Revision D 250 Series Maternal/Fetal Monitor 6-13
2020551-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
5. Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6. Remove card cage cover. Keep screws aside.
7. Disconnect MSpO
2
cable from SpO
2
Carrier board connector.
8. Remove SpO
2
Carrier board with Masimo MSpO
2
Module assembly from
Front-end Motherboard.
9. Insert new SpO
2
Carrier board with Masimo MSpO
2
Module assembly into
Front-End Motherboard.
10. Reconnect MSpO
2
cable to SpO
2
Carrier board connector J10.
11. Replace card cage cover. Re-insert screws.
12. Reconnect 8-conductor Inverter cable to DSP board connector J1.
13. Reconnect 20-conductor DSP-to-LCD Decoder ribbon cable to DSP board
connector J2.
14. Place DSP board onto Main board. Use two screws to fasten DSP board to
chassis.
15. Place top cover. Re-insert screws.
2025177-012 SpO
2
Carrier Board with TruSignal MSpO
2
Module
1. Remove top cover. Keep screws aside.
2. Remove two screws fastening DSP board to chassis.
3. Remove DSP board from Main board.
4. Disconnect 8-conductor Inverter cable from DSP board connector J1.
5. Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6. Remove card cage cover. Keep screws aside.
7. Disconnect MSpO
2
cable from SpO
2
Carrier board connector.
8. Remove SpO
2
Carrier board with TruSignal MSpO
2
Module assembly from
Front-end Motherboard.
9. Insert new SpO
2
Carrier board with TruSignal MSpO
2
Module assembly into
Front-End Motherboard.
10. Reconnect MSpO
2
cable to SpO
2
Carrier board connector J10.
11. Replace card cage cover. Re-insert screws.
12. Reconnect 8-conductor Inverter cable to DSP board connector J1.
13. Reconnect 20-conductor DSP-to-LCD Decoder ribbon cable to DSP board
connector J2.
14. Place DSP board onto Main board. Use two screws to fasten DSP board to
chassis.
15. Place top cover. Re-insert screws.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals