Revision D 250 Series Maternal/Fetal Monitor C-3
2020551-001
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Electromagnetic Compatibility (EMC)
Changes or modifications to this system not expressly approved by GE Medical
Systems can cause EMC issues with this or other equipment. This system is
designed and tested to comply with applicable regulation regarding EMC and must
be installed and put into service according to the EMC information stated in this
appendix.
WARNINGS
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the configuration in which it is being used.
Guidance and Manufacturer’s Declaration – Electromagnetic
Emissions
The Corometrics 250 Series Maternal/Fetal Monitor is intended for use in the
electromagnetic environment specified below. It is the responsibility of the customer
or user to assure that the 250 Series Maternal/Fetal Monitor is used in such an
environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
RF Emissions
EN 55011
Group 1 The equipment uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
EN 55011
Class A
Harmonic Emissions
EN 61000-3-2
Class A The equipment is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Complies

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals