3-8 250 Series Maternal/Fetal Monitor Revision D
2020551-001
Installation: Self-Test Routine
Battery-Backed RAM Status
Whenever you turn off a 250 Series Monitor, a battery provides power to the RAM
(random access memory) that stores information such as time, date, default settings,
etc.
Low Battery Icon
The icon shown above will appear in the upper right-hand section of the monitor
under the following circumstances.
NOTE: It is very important to set the time and date prior to initial operation of the
monitor and during daylight-saving time changes. A long-lasting battery
maintains the settings even when the monitor is unplugged from AC
power. Information on setting the time and date is found in the General
Setup screen section in Chapter 4 of the Corometrics 250 Series Monitor
Operators Manual.
Table 8. Battery-Backed RAM Status
Icon Appearance Reason Solution
Icon appears and then
disappears after power
cycle.
Data corruption. Your monitor has reverted to
factory settings.
Access setup screens and configure last-used
settings.
Icon appears after
multiple power cycles.
Battery requires service. Call GE Service to report.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals