Revision D 250 Series Maternal/Fetal Monitor 4-59
2020551-001
Maintenance: Preventative Maintenance Inspection Report
Page Description Complete Max Actual Pass Fail N/A
Cleaning
4-5 Clean the monitor exterior.
Clean the monitor interior using a hand-held
vacuum.
4-8 Clean the recorder printhead.
4-8 Clean the UA Strain Gauge.
4-6
Clean the monitor accessories (TOCO,
transducers, cables)
4-5 Clean the monitor accessories (cuffs, hoses)
Calibration
Page Description Complete Max Actual Pass Fail N/A
4-53
Verify the power supply voltages on the Main
Board.
4-54
Verify the power supply voltages on the Isolated
Power Supply Board.
5-9 Adjust the printhead.
4-55 Adjust the photosensors on the Recorder Board.
Electrical Safety Tests
Page Description Complete Max Actual Pass Fail N/A
4-10 AC Line (120 V 10%) (<300 mW)
4-10
Ground Impedance (<300 M)
Lead to rear panel ground lug (<300 M)
Lead to top cover bezel screw (<300 M)
(<300 mW)
4-10 Unit to primary leakage
(<300mA)
4-11 Patient-to-Ground leakage for ECG
(
<50mA)
4-12 Patient-to-Line leakage for ECG
(
<50mA)
4-12 Patient-to-Ground leakage for IUP
(
<50mA)
4-13 Patient-to-line leakage for IUP
(
<50mA)
4-13
Patient-to-Ground leakage for MSpO
2
(<50mA)
4-14
Patient-to-Line leakage for MSpO
2
(<50mA)
4-14 Patient-to-Ground leakage for US
(
<50mA)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250 Series and is the answer not in the manual?

GE Corometrics 250 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals