(5) Set the variable supply voltage to 50V ± 0.1V and enter the measured full-scale
voltage:
(6) Press the ENTER key to accept the calibration. Press the EXIT key four times to
quit the CALIBRATION and SETUP modes.
(7) Verify the instrument calibration by applying the test voltages given in Table 2 to
the voltmeter (after both voltage ranges have been calibrated).
(8) Disconnect the calibration/test equipment.
Table 2 Electrical Voltage Input Calibration Tolerances
Voltage Calibration 50V Range Verification Tolerances Voltage Calibration 5V Range Verification Tolerances
Applied Voltage Permissible Deviation Applied Voltage Permissible Deviation
-50 ± 0.05% rdg, ± 0.004% FS -5 ± 0.05% rdg, ± 0.004% FS
-40 ± 0.05% rdg, ± 0.004% FS -4 ± 0.05% rdg, ± 0.004% FS
-30 ± 0.05% rdg, ± 0.004% FS -3 ± 0.05% rdg, ± 0.004% FS
-20 ± 0.05% rdg, ± 0.004% FS -2 ± 0.05% rdg, ± 0.004% FS
-10 ± 0.05% rdg, ± 0.004% FS -1 ± 0.05% rdg, ± 0.004% FS
0 ± 0.05% rdg, ± 0.004% FS 0 ± 0.05% rdg, ± 0.004% FS
10 ± 0.05% rdg, ± 0.004% FS 1 ± 0.05% rdg, ± 0.004% FS
20 ± 0.05% rdg, ± 0.004% FS 2 ± 0.05% rdg, ± 0.004% FS
30 ± 0.05% rdg, ± 0.004% FS 3 ± 0.05% rdg, ± 0.004% FS
40 ± 0.05% rdg, ± 0.004% FS 4 ± 0.05% rdg, ± 0.004% FS
50 ± 0.05% rdg, ± 0.004% FS 5 ± 0.05% rdg, ± 0.004% FS
CALIBRE TENSION
GAMME +/- 50V
Memoriser point PE ?
OUI NON
CALIBRE TENSION
GAMME +/- 50V
RESULTATS
m = 1.395879e-05
c = -7.128004e+00
Applique Mesure
0.0000 526380
49.984 873988
Calibration Terminee
App ENTER pour continuer
CALIBRE TENSION
GAMME +/- 50V
Appliquer PE nominale
App sur ENTER quand
la valeur est stable
REGLER VALEUR
CHANGER
VALEUR
DEMARRER
CALIBRE TENSION
GAMME +/- 50V
Entrer val appliquee:
CHANGER
VALEUR
DEMARRER
CALIBRE TENSION
GAMME +/- 50V
Enter val appliquee:
Calibration