166
Warning
Set Count or
Time of
Condition
Mode of Failure Priority
Probable Cause Criterion for
Clearing Alarm
Corrective Action
Assess patient
Volume
Limit Reached
4 breaths
delay
Set volume limit
has been
reached in
pressure control
modes
Mid Clinical situation
Inspiratory
pressure set too
high
Volume limit is
set too low
Limit volume
(V
T
) is not
reached for 2
breaths
Assess patient
Reduce
Inspiratory
pressure
Increase volume
limit
High PEEP
Alarm
2 breath delay
Actual PEEP
value more than
10 cmH
2
O from
set value
High Exhalation Valve
occlusion
Improper
parameter
setting
PEEP pressure
decrease to set
value + 10 for 1
breath
Check exhalation
valve
Check parameter
setting
Low Minute
Volume
2 breath delay
(20 with
setting
change)
Minute volume
lower than low
alarm setting
High Clinical Condition
Alarm set too
High
Volume is
higher than the
alarm set for
more than 1
breath
Assess patient
Increase alarm
setting if
appropriate
Decrease trigger
sensitivity

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals