7 Care, Maintenance and Tests
175
Cooling Air inlet filter
Replace Cooling Air
inlet filter
Battery
Replace Battery
Ventilator Perform full
calibration
It is good practice to
perform O2
calibration every 3
months.
Perform VVT
Perform safety
checks
Replace listed parts
by an authorized
VersaMed Service
Technician
Every 2 years O2 sensor Replace the O2
sensor
Every 15000 hours or
every 4 years,
whichever is first
Pneumatic Assembly Replace parts by
authorized
VersaMed Service
Technician