Diagnostics/Troubleshooting
7-18 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Network Configuration (continued)
NOTE: If user wants to setup static IP address, uncheck Enable
DHCP option, input static address in IP-Address box,
Subnet Mask and Default Gateway box. In Network
Speed, choose the proper speed available.
Figure 7-11. Input static address
3. Select Save, and a popup window displays. Select OK to
restart the system and activate the changes.
Figure 7-12. System Restart inquiry dialog
4. After the system restarts, the network icon at the left bottom
of screen turns green.
Figure 7-13. Network icon

Table of Contents