Replacement Procedure
LOGIQ e – Basic Service Manual 12-5
5461614-100 English Rev. 6
3-Probe Port Assy (Part No. 5423182-2)
Purpose: This is a description on how to remove and replace the
Docking Cart 3-Probe Port Assy.
Tools
• Common Phillips screwdrivers
• Allen/Unbraco/Torx wrench
Needed Manpower
• 1 person, 1minutes + travel
Preparation
• Shut down the system and switch off the main breaker.