Stress Tests: General Information
The following description is based on the factory default settings. For
instructions on changing the default setup, refer to Chapter 10, “Stress
Test Mode” section.
127(
If you wish to record an ECG only at the end of a test stage, you must
enter a Medians Report time in the protocol editor (Chapter 10,
“Stress Test Mode” section) that is 10 seconds shorter than the stage
time.
Table 1. Setup Menu for the Stress Test Mode
Parameter Factory Default Options
Report Sequence STANDARD CABRERASEQ. No. 3
Protocol WHO see previous page
Rhythm Leads I II, aVF, V2, V4, V6
Gain 10 mm/mV “auto”, 5, 20, 40 mm/mV
Speed 25 mm/s 50 mm/s
In-Test Reports Comparative Median Report Median Report
Muscle Filter No Yes
Muscle Filter Frequency 40 Hz 20 Hz
AC Filter Yes No
max. HR after < 180 WHO 80%, WHO 90%,
WHO 100%, AHA
J+x point formula 60 0, 20, 40, 80, Rautaharju, RR/16
Calculation (E-, J-point) continuous single
Arrhy Data No all, unequal
Final Reports
Summary Report Yes No
Tabular Summary No Yes
Trend Reports No Yes
Trend Reports (ST) No Yes
ST Summary Report No Yes
Episodes no chron., prio, ventr