Revision C MAC 1200/1200 ST 10-33
2012250-021
System Setup: Direct ECG Transmission
Direct ECG Transmission
Select the Automatic mode and press .
Press to display the configuration menu for the Automatic
mode.
Use the cursor keys to position the bar cursor on Manual copy to
HOST and confirm the selection with ([HOST]).
Press to clear the configuration menu.
Use the cursor keys to position the bar cursor on Communication and
confirm the selection with .
Select the same baud rate as at the receiving modem (2400, 4800,
9600, 19200, 38400, 57600).
Selecting the Communication Protocol
This adjustment is only necessary with MAC 1200 / 1200 ST units,
because only these devices can send data to CardioSys/CardioSoft or to
the MUSE system. For a MAC 1100 unit that can send data only to
CardioSys/CardioSoft, the communication protocol A5 is preselected.
1. Using the cursor keys, position the bar cursor on Protocol.
2. Select the protocol A5 if you will send data to CardioSys/CardioSoft
system or select CSI to send data to MUSE system.
3. Use the cursor keys to position the bar cursor on Modem, none and
confirm the selection with .