10-12 MAC 1200/1200 ST Revision C
2012250-021
System Setup: Arrhythmia Mode
Muscle Filter Frequency
Cut-off frequency of the muscle filter: [40 Hz], 20 Hz.
The frequency range is indicated in the lower margin of the recording
strip.
0.08 – 40 Hz (40-Hz muscle filter enabled)
0.08 – 20 Hz (20-Hz muscle filter enabled)
0.08 – 100 Hz (muscle filter off).
Trend Recording
The slow trend recording of 5 mm/s automatically begins at program
start ([no]/yes).
Arrhythmia Data
The recorder will document arrhythmias in the following situations (all
[unequal], no).
Each time an arrhythmia occurs.
Each time an arrhythmia occurs that is different from the preceding
event.
Arrhythmias are not documented.
Episodes
Final report includes episode report, with episodes listed by one of the
following criteria ([chron.], prio., ventr., no.).
In chronological order
According to priorities (see Table 3, “Arrhythmia Codes,” on page 7-9)
Ventricular beats only
No episode report