10-18 MAC 1200/1200 ST Revision C
2012250-021
System Setup: Stress Test Mode
Factory Defaults for Protocols
The factory defaults for treadmill and bicycle ergometer protocols can be
reactivated.
&$87,21
New protocols will be deleted, therefore print your
custom protocols before the reset.
To reactivate factory defaults for treadmill and bicycle ergometer
protocols, position the cursor on the Protocol selection field and press
.
Rhythm Leads
6 leads of your choice for recording of the 6-lead event report.
Gain
When you select *auto the program will automatically determine the
optimal gain setting (*auto, 5, [10], 20, 40 mm/mV).
Speed
Writer speed in mm/s for the printed reports.
In-Test Reports
Format of the In-Test Reports.
Median Report
[Comparative Median Report]
Muscle Filter/AC Filter
Elimination of muscle artifact and AC line interference.
Muscle filter: [no], yes
AC line filter: [yes], no