Automatic Mode Recording: Report Formats
Report Formats
The length and scope of the reports depends on the implemented
software (standard, MEAS (measurement), DIAG (measurement +
interpretation)).
The table below shows all of the 14 different report formats available
with MAC 1200/1200 ST units.
* Measurements and interpretative statements are only available from MAC 1200/
1200 ST with the required software options.
** Only possible with option MEAS or DIAG.
127(
Heart rate is calculated from all beats of the 10-second ECG.
Printed reports are unconfirmed documents. They must be overread,
verified, and signed by a physician for confirmation.
Report formats H1 and A1 are only available with MAC 1200/1200 ST
units (option MEAS or DIAG required). In these reports, the rhythm
leads are printed with a speed of 12.5 mm/s and the medians with
50 mm/s.
Format
ECG Traces
Length/Leads
Rhythm Lead
Length/Leads
Speed Measurement* Interpretation* Pages Medians
12_FS 10 s / 1x12 no 25 mm/s no no 1 no
12_F1 5 s / 1x12 10 s / 1 (V6) 25 mm/s yes yes 1 no
12_F2 8 s / 1x12 10 s / 1 (V6) 50 mm/s yes yes 2 no
6_F1 2x5 s / 2x6 10 s / 1 25 mm/s yes yes 1 no
6_FS 2x5 s/ 2x6 10 s/ 1 25 mm/s no no 1 no
6_F2 2x5 s / 2x6 no 50 mm/s yes yes 2 no
6_F2_25 2x10 s / 2x6 no 25 mm/s yes yes 2 no
3_F1 4x2.5 s / 4x3 10 s / 1 25 mm/s yes yes 1 no
H1** 10 s / 1x6 10 s / 6 12.5 mm/s yes yes 1 12
H2 2x5 s / 2x6 no 50 mm/s yes yes 2 no
A1** 10 s / 1x6 10 s / 6 12.5 mm/s yes + matrix yes 1 12
1x10R3 10 s / 1x3 10 s / 3 25 mm/s yes yes 1 no
4x2.5R3 4x2.5 s / 4x3 3x10 s / 3 25 mm/s yes yes 1 no
MUSE1 4x2.5 s/ 4x3 3x10 s / 2 25 mm/s yes yes 1 no