Revision C MAC 1200/1200 ST 10-19
2012250-021
System Setup: Stress Test Mode
Muscle Filter Frequency
Cut-off filter frequency of the muscle filter (20 Hz, [40 Hz]).
Maximum HR After
For selection of the maximum permitted heart rate:
[<180]
WHO 80%: 180 - patient age
WHO 90%: 200 - patient age
WHO 100%: 220 - patient age
AHA
(age below 25 years: 160,
age from 25 to 75 years: 160 - [(age - 25) x 45/50]
age above 75: 115
J+x Point Formula
0, 20, 40, [60], 80, Rautaharju, RR/16
Calculation (E, J Point)
The E and J point position are either calculated once “single” or
[continuously].
Arrhythmia Data
The recorder will document arrhythmias in the following situations (all,
unequal, [no]).
Each time an arrhythmia occurs.
Each time an arrhythmia occurs that is different from the preceding
event.
Arrhythmias are not documented.