EasyManuals Logo

GE MAC 1200 ST Operator's Manual

GE MAC 1200 ST
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
Revision C MAC 1200/1200 ST 8-17
2012250-021
Stress Tests: Creating Custom Protocols
Creating Custom Protocols
The MAC 1200 ST allows the user to choose among three different test
protocols. The default protocol is the SINGLE protocol with an exercise
phase of 90 seconds. Also available in the list of protocols are the
DOUBLE (180 seconds) and the TRIPLE (270 seconds) protocol.
The protocol can be selected as follows.
1. Press and select Stress test.
2. Press . Now the configuration screen is displayed.
3. Move the cursor to “Protocol [others]” and press : the list of the
available Master's step protocols is shown in the window.
You can also edit the parameters of the post-exercise phase as follows.
1. Press and select Stress test.
2. Press . Now the configuration screen is displayed.
3. Move the cursor to Protocol [SINGLE] and press . The protocol
editor is now active.
The protocol editor can be used to change the number of stages in the
post-exercise phase and the duration of the individual stages. You can
define between 1 and 50 stages. The minimum stage time is 30 seconds.
SINGLE
DOUBLE
TRIPLE

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals