EasyManuals Logo

GE MAC 1200 ST Operator's Manual

GE MAC 1200 ST
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
10-22 MAC 1200/1200 ST Revision C
2012250-021
System Setup: System Setup
Location*
3-place ID number of the location to which the ECG recorder is assigned,
default: 1, range: 1 to 600.
Date/Time
Enter date and time (enter 4 digits for the year).
Lead Fail Beep
Audio signal to indicate electrode problems (yes, [no]).
High HR Beep
Audio signal to indicate that the heart rate exceeds the limit value: yes,
[no] (only in Manual and Arrhythmia modes). The limit value (220 - age)
can be changed manually.
1. The value selected here is the default value appearing in the patient
data
2. with CSI protocol only
Lead Labels
[IEC codes]: R, L, F, N, C1 to C6 or
AAMI codes: RA, LA, RL, LL, V1 to V6
Pace Enhancement
Pacer pulses are displayed enhanced and with a given amplitude ([yes],
no).
When programming pacemakers, we recommend to disable the pace
enhancement.
Baseline Roll Filter
Select the lower cut-off frequency: = 0.04, [0.08], 0.16 Hz.

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals