10-22 MAC 1200/1200 ST Revision C
2012250-021
System Setup: System Setup
Location*
3-place ID number of the location to which the ECG recorder is assigned,
default: 1, range: 1 to 600.
Date/Time
Enter date and time (enter 4 digits for the year).
Lead Fail Beep
Audio signal to indicate electrode problems (yes, [no]).
High HR Beep
Audio signal to indicate that the heart rate exceeds the limit value: yes,
[no] (only in Manual and Arrhythmia modes). The limit value (220 - age)
can be changed manually.
1. The value selected here is the default value appearing in the patient
data
2. with CSI protocol only
Lead Labels
[IEC codes]: R, L, F, N, C1 to C6 or
AAMI codes: RA, LA, RL, LL, V1 to V6
Pace Enhancement
Pacer pulses are displayed enhanced and with a given amplitude ([yes],
no).
When programming pacemakers, we recommend to disable the pace
enhancement.
Baseline Roll Filter
Select the lower cut-off frequency: = 0.04, [0.08], 0.16 Hz.