Revision C MAC 1200/1200 ST 8-11
2012250-021
Stress Tests: Ergometers with Remote Start Pulse
5. When you press or , the program enters the manual mode
and the phase name is given as “Manual”
1
.
For safety reasons, the program maintains the current stage and the
buttons and are inactive. You can only control the test
manually with the buttons .
6. When you press , the program increments to the next stage and
a stage report is printed. With you advance to the next phase.
1
With the EC1200, takes you to the next higher stage and
to the preceding stage without generating a stage report. After a load
change, wait until the new load appears in the display before you
activate one of these buttons again.
When the number of revolutions is low, the EC1200 will
automatically activate a load of 15 Watts and increase the load to the
old value when the speed picks up again. During that time, the
buttons , , are inactive and the program will not
proceed to the next load stage.
1
Does NOT apply to the EC1200.
Alt