Revision C MAC 1200/1200 ST 5-15
2012250-021
Automatic Mode Recording: ECG Transmission
127(
If the MAC 1200/1200 ST unit is configured for manual dialing,
the prompt Please Dial the number... will be display. Enter the
number to which the ECG is being transmitted. See “Manual
Dialing” on page 10-27 for information on configuring the
manual dialing option.
6. As soon as you initiate the transmission with , the unit
automatically dials the number of the modem at the receiving end
and establishes a connection.
The message Initializing transmission is displayed, and the option to
cancel the transmission is provided. Then it will send the ECG.
7. After the transmission, a message on the display indicates the
number of successfully transmitted ECGs. As soon as you
acknowledge the message with , the Automatic mode acquisition
screen appears.
8. If the ECG could not be transmitted (wrong modem setup, modem
off, etc.), the unit will display an error message, such as
Transmission Error! (A5).
In this situation you have the following choices:
Repeat the transmission with .
Change the settings with .
Stop the transmission with .
You must select one of these choices to clear the transmission error
message from the display.
127(
Unsuccessful transmission does not remove/delete the ECG
record from the local database.