Arrhythmia Mode Recording: Recording
The locations of all relevant device settings are shown below:
Refer to Table 3, “Arrhythmia Codes,” on page 7-9 for an explanation of
the arrhythmia codes annotated on the recording.
The heart rate limit is automatically calculated from the date of birth
(WHO 100% = 220 - age). When the date of birth is not entered, the unit
will set the limit at 180 bpm. This value can be changed with and
(in steps of 5 bpm).
127(
The letter A on a recording indicates the presence of artifact which
does not allow the algorithm to identify arrhythmias. Causes include
wandering baselines. The anti-drift system largely prevents these
disturbances. Check the electrodes and leadwires.
Operating mode Muscle filter
1
1. if enabled
AC line filter
1
Anti-drift system
1
“Patient Name”
Report function
2
2. or REC OFF when the recording function is disabled
Gain Report sequence
Three leads
display here
Failure message
3
3. if any
Heart rate

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals