Chapter 8. Maintenance and Troubleshooting
Model PM880 Portable Hygrometer User’s Manual 155
Err 11:
No Reference
No reference values were 
entered for a moisture or 
pressure probe.
Enter reference values.
Err 12: No 
Calibration
No calibration curve was entered 
for a moisture or pressure probe.
Enter calibration curve.
Err 15:
Cal Error
An internal calibration 
reference has failed during 
AutoCal, or a probe failure is 
interfering with AutoCal.
Contact the factory.
Err 25:
No Data
The MIS Probe has not 
responded, or a calculation is not 
yet complete.
The PM880 will try to send data; 
wait for next update.
MIS Probe
Calibration 
data will be 
written to the 
attached MIS 
probe.
This screen appears when new 
calibration data is manually 
entered in the Calibrate menu.
Press 
[F3] (OK) to overwrite 
existing data. Pressing OK 
overwrites the calibration data 
stored in the MIS probe electronics 
module.
Press [F2] (Cancel) to 
discontinue.
These setup 
values will be 
written to the 
attached MIS 
probe.
This screen appears when new 
setup data is manually entered in 
the Service, MIS Probe menu.
Press 
[F3] (OK) to 
overwrite existing data. Pressing 
OK overwrites the setup data (e.g. 
reference values) stored in the MIS 
probe electronics module.
Press 
[F2] (Cancel) to 
discontinue.
Table 5: Error and Screen Messages (cont.)
Error 
Message Problem Action