Chapter 4. Using Special Features
50 Model PM880 Portable Hygrometer User’s Manual
4.5 Setting a Probe Calibration Reminder
The Probe ID command enables you to enter the probe serial number,
calibration dates and a calibration reminder. Use the following steps to enter
probe data:
4.5.1 Accessing the Probe ID Tab
1. If the menu is not active, press [MENU].
2. Use the arrow key to scroll to Program and press [ENTER].
3. Use the arrow key to scroll to Calibrate and press [ENTER].
4. Use the arrow keys to scroll to the Probe ID tab and press [ENTER]. A
screen similar to Figure 20 appears.
Figure 20: Probe ID Window