4.6 Using Automatic Calibration (AutoCal)
The PM880 automatically calibrates the moisture and pressure
measurement circuitry (AutoCal) at a user-selected interval. Autocal
compensates for any drift in the electronics. GE recommends setting the
AutoCal interval to 8 hours. Set a smaller AutoCal interval more frequently
if the meter is exposed to extreme temperatures or weather conditions.
Note: If you are using only Moisture Image Series (MIS) probes to measure
moisture, the System Menu is not available.
Use the steps below to enter an AutoCal interval:
1. If the menu is not active, press [MENU].
2. Use the arrow key to scroll to Program and press [ENTER].
3. Use the arrow key to scroll to System and press [ENTER]. A screen
similar to Figure 21 appears.
Figure 21: System Configuration Window