Chapter 8. Maintenance and Troubleshooting
Model PM880 Portable Hygrometer User’s Manual 161
8.14Recalibrating the Pressure Sensors
Since the pressure sensor on a TF Probe or on Moisture Image Series (MIS)
Probes is a strain gage type, the pressure calibration is linear and is
calibrated at 2 data points. Each point consists of a voltage or FP value and
a corresponding pressure value in psig. Check or change the two calibration
data points using the steps below.
1. Set one of the boxes on the screen to display pressure in mV (or FP) and
another box to display psig. Refer to Displaying Measurements on
page 15 to set up the screen.
2. Expose the pressure sensor to air and record the mV (or FP) and psig
readings. This reading is for the zero pressure.
3. Expose the pressure sensor to a known full scale pressure source and
record the mV (or FP) and psig readings. This reading is for the span
pressure.
4. Press [MENU].
5. Use the arrow key to scroll to Program and press [ENTER].
6. Use the arrow key to scroll to Calibration and press [ENTER].
7. Proceed to Entering Pressure Calibration Data on page 14 to complete
entering data.

Table of Contents

Other manuals for GE PM880

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PM880 and is the answer not in the manual?

GE PM880 Specifications

General IconGeneral
BrandGE
ModelPM880
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals