Section 8 — Send and Receive Information
155
Resend Updated Folder
You will usually create all your machines in Ascent software and send
these to your instrument only once. However, if you add or change a
route, machine, or parameter set in the software, you will need to resend
that folder to the instrument so it can update to include your changes.
Each time you resend an existing folder from Ascent to your instrument,
it will overwrite the contents of the folder on the instrument. Any
machines, routes, and parameter sets, etc that you have added or
deleted in Ascent will also be added or deleted on the instrument.
Additionally, all stored recordings in the instrument are also deleted. This
is designed to free instrument memory and save you the task of
manually deleting recordings, machine definitions, and more.
Retain Stored Recordings
If you do NOT wish to overwrite an existing recording, simply rename the
folder on your instrument or in Ascent before you resend it. Your
instrument will treat this as a new folder and will NOT overwrite the
contents of the existing folder.
Retain Baseline Recordings on Instrument
If a recording is marked as Baseline in Ascent, it will be transferred to the
instrument when the folder is resent. If a different recording is marked as
baseline in the instrument, this will be replaced with the newly sent
baseline.