Section  4  ā  Measurement  Types  and  their  Uses 
72 
2 ā Setup  
 
Before  you  can  use  your  instrument  to  take  a  Cross  Channel  Spectrum 
(ODS)  recording,  you  will  need  to  specify  the  appropriate  settings  for  your 
channels  and  sensors.  You  will  need  at  least  two  sensors  enabled  for  this 
type  of  recording.  The  lowest  channel  (e.g.  CH1)  will  be  used  as  the 
reference  channel.  
 
To  set  up  your  channels  and  sensors:   
1.  Press    or    Measure.   
2.  Use  the  navigation  arrows  to  select  Cross  Channel  Spectrum  (ODS) 
>  Press  . 
3.  Select  your  preferred  FRF  unit  ā  Accelerance  /  Mobility  / 
Compliance: 
a.  Press    Store  Units. 
b.  Use  the  navigation  arrows  to  select  your  preferred  unit  >  Press 
. 
 
NOTE: This selection is the modal equivalent of choosing between 
acceleration, velocity, and displacement. Consequently, 
Accelerance is the better choice for very high frequencies, while 
compliance is the better option for very low frequencies. 
Frequency Response Function (FRF) units are derived automatically 
based on your unit selection.  
 
4.  Specify  your  preferred  spectral  lines  and  samples: 
a.  Press    Spectrum  Waveform. 
b.  Use  the  navigation  arrows  to  select  the  required  option  > Press 
.