Section 4 Measurement Types and their Uses
73
5. Select the required Fmax setting Ensure this value covers any
frequencies at which the machine could exhibit mechanical
resonance:
a. Press Fmax Fmin.
b. Use the navigation arrows to select the required option > Press
.
NOTE: If required you can press to disable automatic
incrementing of the reference sensor (lowest channel number) as
you take your recordings (NOT recommended).
6. Press to access additional setup options:
a. Press Number of Averages to change the default number
of averages taken.
b. Press Point Name Increment to change the default
interval at which points will automatically increment.
c. Press Average Overlap repeatedly to toggle and select
the required percentage.
d. Press Window to toggle between Hanning and
Rectangular options.
e. Press Sensor Setup to set up your channels and sensors
as required.
7. Press to exit when you have finished selecting the required
parameters You can now start recording.