DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
4-22 Chapter 4 - General Procedures and Functional Checks
PRELIMINARY
4-3-6-2 Cart Mechanical Checks
4-3-6-2-1 Cart Up/Down Movement Test
1.) While the system is mounted on the cart, push down the pedal located at the base of the cart
downwards and verify the system may be lowered and raised easily
Figure 4-58 System UP/DOWN Lever

Table of Contents